Quick summary: Alinity M Resp-4-Plex Amplification (Amp) Kit (Ce). Date: 2022-12-14.
Alinity M Resp-4-Plex Amplification (Amp) Kit (Ce)
2022-12-14
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Alinity M Resp-4-Plex Amplification (Amp) Kit (Ce). Date: 2022-12-14.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
There is a potential for false positive results for flu B and RSV while using the list numbers and lot numbers referenced above associated with the manufacture of these lots. There is also a potential for delayed results when using these lots as reactive negative controls invalidate the sample run. Additional quality control mitigations have been put in place to prevent reoccurrence. All subsequent lots of Alinity m resp-4-plex amp kits are not impacted. Recall start date: November 23, 2022
What you should do
- Contact the manufacturer fi you require additional information.