Quick summary: Alinity HQ Analyzer. Date: 2024-07-05.
Alinity HQ Analyzer
2024-07-05
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Alinity HQ Analyzer. Date: 2024-07-05.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Abbott has identified the following issues when using Alinity HQ Analyzer software version 5.6 and below which will be corrected in software version 5.9. Issue 1: When processing a CBC+diff sample that immediately follows a CBC+diff+retic sample, diluent may backflow into the RBC dilution cup. This may result in falsely low red blood cell (RBC) results and consequently falsely high mean cell hemoglobin concentration (MCHC), falsely high mean cell hemoglobin (MCH), and falsely low hematocrit (HCT). Issue 2: An overestimation of the basophil (baso) count may occur on some samples when cell events are incorrectly counted as baso during sample analysis. this may result in falsely increased baso counts and %b (%baso). Recall start date: June 21, 2024
What you should do
- Contact the manufacturer if you require additional information.