Quick summary: Alinity C Iron Reagent Kit. Date: 2023-10-16.
Alinity C Iron Reagent Kit
2023-10-16
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Alinity C Iron Reagent Kit. Date: 2023-10-16.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Sentinel Ch Spa, the manufacturer of iron for use with Architect and Alinity C Iron Reagent Kit, has identified an issue with the reagents. Some customers have experienced elevated quality control (QC) results as well as elevated patient results. A high recovery of iron assay has been observed at some customer sites. Internal studies confirmed the positive shift of quality control and patient samples results, up to 15% for iron products (Alinity C Iron Reagent ref. 08p3920) to be used on Abbott Alinity instrument and 30% for iron products (iron ref. 6k95-41 and iron ref. 6k95-30) to be used on Abbott Architect instrument. Recall start date: September 29, 2023
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.