Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite™ Y-site(s)
2024-05-03
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) Back Check Valve 2 SmartSite™ Y-site(s). Date: 2024-05-03.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
BD has received customer reports of the Alaris™ Pump Infusion Set Low Sorbing Tubing (PE Lined) drip chamber becoming detached from the tubing. BD has confirmed that this issue is isolated to catalog number 10013072 only and a single batch (no. 22115368) where mis-assembly of the drip chamber to the tubing occurred. All other active catalog numbers containing a drip chamber are not affected. Recall start date: April 18, 2024
What you should do
- Contact Becton Dickinson Canada Inc.