Quick summary: A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. Reason: Health product recall. Category: Health products. Date: 2026-06-23.
A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors
2026-06-23
Recommended action: compare your product with the official notice and follow the instructions published by the official source.
Recall details
Quick summary: A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. Reason: Health product recall. Category: Health products. Date: 2026-06-23.
Affected product details
Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.
Affected products
Issue
Zimmer Biomet is conducting a medical device recall for A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. One (1) complaint was received with no clinical impact reporting that wrinkles were identified in the sealing area of the sterile pouch. Subsequent seal integrity testing conducted internally showed potential sterility breach in 0.92% of inspected units that had visual seal nonconformities. The devices are distributed in cases (CS) of 20 units. Inside each case are two boxes (BX) of 10 units each.
What you should do
- Contact the manufacturer if you require additional information.
Additional official information
Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.
Frequently asked questions
Answers based on the official recall notice.
Product images
Official product images from Health Canada.
Starting...
Explore more
Related recalls
Similar recalls and safety notices.