Public recall detail

A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors

2026-06-23

medium
Health products Health product recall
Quick summary

Quick summary: A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. Reason: Health product recall. Category: Health products. Date: 2026-06-23.

Recommended action

Recommended action: compare your product with the official notice and follow the instructions published by the official source.

Recall details

Quick summary: A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. Reason: Health product recall. Category: Health products. Date: 2026-06-23.

Official details

Affected product details

Structured information extracted from the official recall notice. Confirm the identifiers and instructions in the original notice.

BrandZimmer Surgical, Inc. Also Trading As Relign Corporation
ProductA.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors
IssueMedical devices - Sterility
Recall classType II
IdentificationRA-82244

Affected products

ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot79701948
Model60-4009-101-00
Serial number79701948
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80555938
Model60-4009-101-00
Serial number80555938
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot79701947
Model60-4009-101-00
Serial number79701947
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80581870
Model60-4009-101-00
Serial number80581870
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80765021
Model60-4009-101-00
Serial number80765021
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot79701946
Model60-4009-101-00
Serial number79701946
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80555940
Model60-4009-101-00
Serial number80555940
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80555939
Model60-4009-101-00
Serial number80555939
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
Lot80581869
Model60-4009-101-00
Serial number80581869
DescriptionA.T.S® 4000 TS Tourniquet Systems Single Hose With CPC Connectors
ProductA.T.S® 4000 TS Tourniquet Systems Dual Hose With CPC Connectors, 20-Pack
LotMore than 10 numbers, contact manufacturer.
Model60-1812-101-00
Serial numberMore than 10 numbers, contact manufacturer.
DescriptionA.T.S® 4000 TS Tourniquet Systems Dual Hose With CPC Connectors, 20-Pack

Issue

Zimmer Biomet is conducting a medical device recall for A.T.S® 4000 TS Tourniquet Systems Single and Dual Hose with CPC Connectors. One (1) complaint was received with no clinical impact reporting that wrinkles were identified in the sealing area of the sterile pouch. Subsequent seal integrity testing conducted internally showed potential sterility breach in 0.92% of inspected units that had visual seal nonconformities. The devices are distributed in cases (CS) of 20 units. Inside each case are two boxes (BX) of 10 units each.

What you should do

  • Contact the manufacturer if you require additional information.
Additional official information
AudienceHealthcare
CategoryHealth products - Medical devices - General and plastic surgery
CompaniesZimmer Surgical, Inc. Also Trading As Relign Corporation200 W Ohio Avenue, Dover, Ohio, United States, 44622
Recall date2026-06-15
Published byHealth Canada
Recall classType II
Alert / recall typeHealth product recall
Identification numberRA-82244
Background{}
CompaniesZimmer Surgical, Inc. Also Trading As Relign Corporation200 W Ohio Avenue, Dover, Ohio, United States, 44622
AudiencesHealthcare
Source updated2026-06-23

Contains information licensed under the Open Government Licence – Canada. SafetyPulse is independent of the Government of Canada.